| Data Summary | |||||||||||
| Optimal Accuracy Criteria | Sensor | Meter | |||||||||
| Date | Number of Paired Sensor/Meter Readings | Correlation Coefficient [r] | Mean Absolute Error (%) | Number of Readings | Average (mg/dL) | SD (mg/dL) | Sensor Range (mg/dL) | Number of Readings | Average (mg/dL) | SD (mg/dL) | Meter Range (mg/dL) |
| 06-äæÝãÈÑ-02 | 3 | n/a | 16.7 | 141 | 109.2 | 22.9 | 49-149 | 3 | 73.3 | 13.7 | 55-88 |
| 07-äæÝãÈÑ-02 | 3 | 0.98 | 16.2 | 252 | 102.7 | 14.0 | 77-141 | 6 | 223.7 | 125.9 | 104-400 |
| 08-äæÝãÈÑ-02 x | 1 | n/a | 16.5 | 97 | 104.7 | 6.7 | 86-117 | 3 | 251.7 | 110.8 | 95-334 |
| All Days | 7 | 0.85 | 16.5 | 490 | 105.0 | 16.4 | 49-149 | 12 | 193.1 | 126.3 | 55-400 |
| x: This day does not satisfy the criteria for optimal accuracy as indicated by the shaded entries in the summary table. | |||||||||||
| Please use your clinical judgment in evaluating the graph. | |||||||||||
| c: This day has no paired sensor/meter data. As a result, only meter data is provided. | |||||||||||
| s: This day has no meter data. As a result, only sensor profile is provided. | |||||||||||
| Note: If the Meter Range difference between high and low values for a day's range is less than 100 mg/dL or 5.6 mmol/L, | |||||||||||
| then the Correlation Coefficient for that day will be reported as 'n/a'. | |||||||||||