| Data Summary | |||||||||||
| Optimal Accuracy Criteria | Sensor | Meter | |||||||||
| Date | Number of Paired Sensor/Meter Readings | Correlation Coefficient [r] | Mean Absolute Error (%) | Number of Readings | Average (mg/dL) | SD (mg/dL) | Sensor Range (mg/dL) | Number of Readings | Average (mg/dL) | SD (mg/dL) | Meter Range (mg/dL) |
| 29-اكتوبر-02 | 3 | n/a | 9.7 | 141 | 230.9 | 37.0 | 165-295 | 3 | 253.7 | 19.7 | 228-276 |
| 30-اكتوبر-02 x | 3 | n/a | 20.2 | 288 | 233.5 | 55.1 | 81-348 | 3 | 238.7 | 28.5 | 217-279 |
| 31-اكتوبر-02 x | 2 | n/a | 4.4 | 116 | 179.2 | 46.1 | 119-257 | 10 | 218.8 | 20.5 | 176-246 |
| All Days | 8 | 0.13 | 12.3 | 545 | 221.3 | 53.8 | 81-348 | 16 | 229.1 | 26.2 | 176-279 |
| x: This day does not satisfy the criteria for optimal accuracy as indicated by the shaded entries in the summary table. | |||||||||||
| Please use your clinical judgment in evaluating the graph. | |||||||||||
| c: This day has no paired sensor/meter data. As a result, only meter data is provided. | |||||||||||
| s: This day has no meter data. As a result, only sensor profile is provided. | |||||||||||
| Note: If the Meter Range difference between high and low values for a day's range is less than 100 mg/dL or 5.6 mmol/L, | |||||||||||
| then the Correlation Coefficient for that day will be reported as 'n/a'. | |||||||||||