| Data Summary | |||||||||||
| Optimal Accuracy Criteria | Sensor | Meter | |||||||||
| Date | Number of Paired Sensor/Meter Readings | Correlation Coefficient [r] | Mean Absolute Error (%) | Number of Readings | Average (mg/dL) | SD (mg/dL) | Sensor Range (mg/dL) | Number of Readings | Average (mg/dL) | SD (mg/dL) | Meter Range (mg/dL) |
| 15-äæÝãÈÑ-02 | 3 | 0.99 | 20.2 | 142 | 139.0 | 46.7 | 70-213 | 3 | 116.3 | 51.2 | 62-185 |
| 16-äæÝãÈÑ-02 x | 4 | -0.93 | 33.1 | 288 | 209.2 | 64.7 | 86-400 | 4 | 265.3 | 52.8 | 212-320 |
| 17-äæÝãÈÑ-02 x | 0 | n/a | n/a | 105 | 167.4 | 86.1 | 81-400 | 66 | 350.4 | 32.2 | 294-400 |
| All Days | 7 | 0.53 | 27.6 | 535 | 182.4 | 72.2 | 70-400 | 73 | 336.1 | 60.4 | 62-400 |
| x: This day does not satisfy the criteria for optimal accuracy as indicated by the shaded entries in the summary table. | |||||||||||
| Please use your clinical judgment in evaluating the graph. | |||||||||||
| c: This day has no paired sensor/meter data. As a result, only meter data is provided. | |||||||||||
| s: This day has no meter data. As a result, only sensor profile is provided. | |||||||||||
| Note: If the Meter Range difference between high and low values for a day's range is less than 100 mg/dL or 5.6 mmol/L, | |||||||||||
| then the Correlation Coefficient for that day will be reported as 'n/a'. | |||||||||||